YX-DL8500 Automatic sulfur analyzer

Instrument standard configuration: 1 computer + 1 printer + 1 set of YX-DL8500

INQUIRY
  • Specification
  • Competitive Advantage

Scope of application
Suitable for power plants, coal mines, commodity inspection, environmental protection, geological exploration, metallurgy, papermaking, chemical industry, scientific research, education and other departments to determine the total sulfur content of coal, coke, petroleum, solid hazardous waste and other substances.

Technical parameter
Measuring range: 0.01%~40%
Test method: Coulometric titration
Sulfur measurement resolution: 0.001%
Single sample analysis time: about 4min
Working furnace temperature: 1150°C (coal, coke), 900°C (oil)
Temperature control accuracy: ±2℃
Sample mass: 50±5mg (coal, coke), 100±10mg (petroleum products)
Sample delivery method: automatic sample delivery, sample abandonment
Number of samples: 19/batch, samples can be added/replaced during the test
Accuracy: in line with GB/T214-2007 "Determination of Total Sulfur in Coal"
Working power supply: AC220V±22V/50Hz
Power: ≤3kW
Dimensions (mm): 1010×500×380
Mass of the whole machine (kg): 78

Technical Features
● Adopting a horizontal integrated structure, the automatic sample feeding, sample feeding mechanism, high temperature cracking furnace, electrolytic cell, agitator, air purification system and other components are skillfully combined, making the instrument structure more compact, more reasonable and more stable;
● Using a special flat-bottom corundum porcelain boat to reduce the possibility of jamming. Program control realizes automatic heating, temperature control, sample delivery, and measurement, and the results are automatically saved and printed, and samples can be added in the middle, which greatly improves work efficiency and truly realizes unattended operation;
● Coulometric titration method is used for measurement, and the measurement time can be automatically judged according to different samples, and the speed is better than that of Ashka gravimetry and high temperature combustion neutralization method;
● Adopt linear electrolyte molecular concentration acquisition circuit to effectively solve the problem of over-electrolysis;
● Adopt non-contact switch, with burnout protection, hardware over-temperature protection and electrolyte backflow prevention measures;
● The multi-point dynamic coefficient correction method is adopted to ensure higher accuracy of the test results of coal with high, medium and low sulfur content;
● It can be connected to the laboratory management system, so that the balance data     does not fall to the ground, and the test results are backed up and uploaded;
● It can be used to determine the total sulfur content in samples such as solid hazardous wastes.