ASTM D93 FDT-0201, Open cup flash point and fire point tester
The instrument is designed and manufactured in accordance with ASTM D93 standard.
INQUIRY- Specification
- Competitive Advantage
Place of Origin: | Changsha,China |
Brand Name: | FRIEND |
Model Number: | FDT-0201 |
Certification: | CE, ISO,CO |
Minimum Order Quantity: | 1 SET |
Price: | USD600-2000 |
Packaging Details: | Hard carton,Standard export packing |
Delivery Time:(LEADING) | 30 DAYS |
Payment Terms: | T/T, L/C |
Supply Ability: | 50SET/MONTH |
Applications:
1. ASTM D93
2. Gasoline, lubricant oil, fuel oil testing
3. Petroleum, power, chemical, research, quality control and university
Specifications:
Technical Parameters | |
Power supply | AC 220V ± 10%, 50Hz |
The heating device | Heating power from 0 ~ 600W continuously adjustable |
Sample heating rate | 1 ~ 12 ℃ / min, it can be controlled. |
Electric mixing device | (1) 90 ~ 120 r / min for the test step A. (2) 250 ± 10 rev / min for the test step B.
|
Electric mixing device | (1) 90 ~ 120 r / min for the test step A. (2) 250 ± 10 rev / min for the test step B.
|
Standard oil cup | (1) Diameter: 50.7 ~ 50.8 mm. (2) Depth: 55.7 ~ 56.0 mm. (3) Test oil capacity groove depth: 33.9 ~ 34.3 mm. (4) Test oil capacity: about 70 ml.
|
Thermometer | A bar mercury thermometer, in line with GB / T 261-2008 standards. (1) temperature range -5 ~ 110 ℃, indexing 0.5 ℃. (2) temperature range of 20 ~ 150 ℃, indexing 1.0 ℃. (3) temperature range of 90 ~ 370 ℃, dividing 2 ℃.
|
Ignition | (1) ignition sources: gas (or other civilian combustible gas, the same below). (2) pilot aperture: 0.8 mm. |
Ambient temperature: | ≤35 ℃. |
Relative humidity | ≤85%. |
Total power consumption | Less than 650W. |
Feature | 1. the instrument is designed in full compliance with ASTM D93, GB/T261 standards. 2. heating power continuously, stepless adjustable, voltmeter intuitive display power size, advanced, reasonable temperature control. 3. compact structure, stainless steel countertops, easy to clean oil. 4. easy to operate, the test results are accurate, the instrument was a number of measurement units selected as closed flash point detection equipment. |