ASTM D6079, FDR-5471 Dizal lubric tester(HFRR)
Complies with SH/T0765-2005 "Diesel Lubricity Assessment Method (High Frequency Reciprocating Tester Method)", ASTM D6079 "Test Method for Evaluating Diesel Lubricity with High Frequency Reciprocating Test Bench (HFRR)" (ISO12156-1:1997, MOD standard method) required technical indicators.
BINCIKE- Ƙayyadaddun bayanai
- Amfanin da ya dace
Aikace-aikace:
1. ASTM D6079,SH/T0765
2. Diesel
3. Man fetur, wutar lantarki, sinadarai, bincike, kula da inganci da jami'a
bayani dalla-dalla:
Technical sigogi | |
Maximum rated displacement (stroke) of the vibration exciter | ±3mm; accuracy 0.001mm
|
Experimental temperature | Room temperature -200 ° C, any adjustment |
Daidaitawar yanayin zafin jiki | ± 0.1 ° C |
Gwajin ƙarfin gwaji | 0-1Kg, friction measurement range: 0-45N |
Girma: | 100 times; accuracy: 1 micron; |
Constant temperature and humidity box:
| Temperature control range: 17~28°C, temperature display accuracy 0.1°C, temperature control accuracy ±2°C |
Control humidity range:
| 25-70%, humidity display accuracy 1%, humidity control accuracy ≤ 3% RH; |
Yanayin aiki:
| Temperature 5 to 35 ° C, humidity: 20-80% |
Working power: , | AC220V, 50HZ |
The whole machine power: | 2000W |
Reciprocating displacement | mm 1.0 ± 0.02 |
Reciprocating frequency /Hz | 50 ± 1 |
Application load / g | 200 ± 1 |
Oil sample volume / ml | 2 ± 0.2 |
Oil sample temperature / °C | 60 ± 2 |
Test time / min | 75 ± 0.1 |
| |
Features | 1. Wear scar measurement system: electronic imaging system, can collect and store images; x, y two directions of reading; magnification: 100 times; accuracy: 1 micron
|