YX-GYFX7706D Automatic industrial analyzer

YX-GYFX7706D Automatic industrial analyzer

Instrument standard configuration: 1 computer + 1 printer + 1 set of YX-GYFX7706D

Ëmfro
  • bestëmmen
  • Kompetitiv Virdeel

Ëmfang vun der Applikatioun
It is used in power plants, coal mines, commodity inspection, environmental protection, chemical industry, scientific research, education and other departments to determine the moisture, ash and volatile content of coal, coke, petroleum and other substances, and automatically calculate the calorific value, fixed carbon and hydrogen content.

Technesche Parameter
Temperature control range: RT+10~1000℃
Temperatur Kontroll Genauegkeet: ± 2 ℃
Sample mass: 0.5~1.2g, recommended: 1±0.1g
Number of samples: 20/batch
Test time: 240min/20 samples
Aarbecht Muecht Fourniture: AC 220V ± 22V / 50Hz
Leeschtung: ≤4 kW
Dimensiounen (mm): 630 × 560 × 580
Gewiicht (kg): 90

Technesch Features
● Adopt MCPC control module and balance display module to monitor the status of the instrument in real time, save the instrument parameters, complete the test process independently, without the intervention of the host computer, realize offline operation, and self-diagnosis reminder of the instrument, which greatly improves the reliability of the instrument;
● Benotzt Steppermotor a High-Präzisioun Wénkel Encoder Kontrollsystem (Utility Model Patentnummer: 2017205786217), präzis Positionéierung a séier Probe waacht;
● Mat der Power-Off Memory Funktioun ginn d'Testdaten net verluer wann den Testprozess onerwaart ënnerbrach gëtt, an den Test kann automatesch ofgeschloss ginn wann d'Muecht ageschalt ass;
● The use of distributed intelligent smoke exhaust interface can more effectively exhaust smoke and dust, and is more environmentally friendly;
● Using PID temperature control algorithm, the temperature control accuracy is higher;
● Adopt automatic flip device to avoid high temperature radiation and burns of operators;
● Using thermogravimetric analysis method, automatic sample weighing, automatic sample delivery, automatic data processing, result calculation, report printing and storage, etc., to achieve unattended;